Flow-Through Microvial Facilitating Interface of Capillary Isoelectric Focusing and Electrospray Ionization Mass Spectrometry

TitleFlow-Through Microvial Facilitating Interface of Capillary Isoelectric Focusing and Electrospray Ionization Mass Spectrometry
Publication TypeJournal Article
Year of Publication2011
AuthorsZhong, X, E. Maxwell, J, Ratnayake, C, Mack, S, Chen, DDY
JournalANALYTICAL CHEMISTRY
Volume83
Pagination8748-8755
Date PublishedNOV 15
ISSN0003-2700
Abstract

A capillary electrophoresis mass spectrometry (CE-MS) interface utilizing a flow-through microvial is used to ensure the electric continuity and supply the catholyte and mobilizer solutions during the capillary isoelectric focusing (cIEF) and mobilization process. The flow-through microvial provides a stable chemical environment and helps to improve the ionization efficiency without significantly diluting the analyte. The CE-MS interface facilitates the transfer of the mobilized cIEF effluent to the site of electrospray ionization, and the gaseous ions can be detected directly by a mass spectrometer. It also allows for complete focusing and mobilization processes to be performed automatically in programmed sequences with commercial CE systems. Two different strategies, using either a part of the capillary or the flow-through microvial of the CE-MS interface as the catholyte reservoir for bare fused silica capillaries or neutral coated capillaries, respectively, were developed for automated cIEF-electrospray ionization (ESI)-MS. Reasonable separation efficiency was achieved using proper concentration of carrier ampholytes and suitable strategies of electroosmotic/electrophoretic mobilization.

DOI10.1021/ac202130f